My pump P2.1S/4.1S shows a flow rate in analog mode, but does not start pumping. What's wrong?

When controlling the pump via analog in (e.g. with a flow meter), the eluent delivery must be started using the start-in connector. Simply short-circuit START-IN and GROUND and press the Start-stop key to start delivery.