The AZURA UV/VIS Detector UVD 2.1L is a competitively priced HPLC spectrophotometer for routine HPLC applications including fast LC methods. Besides offering excellent technical specifications, this robust detector features a highly flexible and compact design. The UVD 2.1L comes with an installed deuterium lamp which covers a wavelength range from 190 to 750 nm. This detector can be controlled with OpenLab® EZChrom, ChromGate®, and ClarityChrom® software, as well as with the Mobile Control software (stand-alone operation), via LAN or through analog input/output; allowing it to be integrated into almost any LC system. Due to a smart design the flow cell is easily accessible and can be changed very quickly. You can choose between a wide range of flow cells for analytical or preparative LC applications with flow rates from 10 μl/min up to 10 l/min.

Key Features

Technical Data


Detector type Variable single wavelength UV detector
Detection channels 1
Light source Deuterium (D²) lamp with integrated GLP chip
Wavelength range 190 - 750 nm
Spectral bandwidth 11 nm at Hα line (FWHM)
Wavelength accuracy ± 2.5 nm
Wavelength precision 0.3 nm (ASTM E275-93)
Noise ± 1.5 x 10-5 AU at 254 nm (ASTM E1657-98)
Drift 3.0 x 10-4 AU/h at 254 nm (ASTM E1657-98)
Linearity > 2.0 AU at 270 nm (ASTM E1657-98)
Maximum data rate 50 Hz (LAN)/20 Hz (Analog)
Flow cell Not included (see Accessories / Spare parts)
Time constants 0.0 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 / 5.0 / 10.0 s
Integration time Automatic
Leak sensor Yes


Inputs Error (IN), Start (IN), Autozero, 0 - 10 V Analog IN
Outputs Events 1 - 3, + 5 V, 24 V Valve
Analog outputs 1 x 0 - 5 V scalable, 20 bit, offset adjustable
Control Digital: LAN, remote connector/Analog: wavelength control/Manual: Mobile Control (optional)
Programming Timed: wavelength, events, fraction valve, links, wake up (program, link); /9 programs, 50 program lines

Technical parameters

GLP Detailed report including lamp recognition, operating hours, /lamp operating hours, number of lamp ignitions
Display Mobile Control (optional)
Ambient conditions Temperature range: 4 - 40 °C, 39.2 - 104 °F, Humidity: below 90 % noncondensing


Power supply 100 - 240 V, 50 - 60 Hz, 75 W
Dimensions 361 x 158 x 523 mm (W x H x D)
Weight 5.9 kg


Note Flow cells are not included and must be ordered separately (see Accessories / Spare parts)