Technical data are subject to change without notice.
All trade names and registered trademarks are the property of their respective owners.
Dissolved gases in the solvent can cause bubbles in the flow path of pumps and detectors. Reliable chromatographic separation therefore requires degassing of the solvent. The small analytical 2-channel degasser DG 2.1S is equipped with two degassing chambers and can thus degas two solvents simultaneously.
Degasser channels | 2 |
---|---|
Max. flow rate/channel | 10 ml/min |
Recommended flow rate/channel | 2.8 ml/min |
Degassing method | Gas permeation through a fluoropolymere membrane |
Degassing chamber volume | 285 µl |
Solvent applicability | universal, except hydrochloric acid, halogenated hydrocarbons, hexafluoro isopropanol (HFIP), hexane, heptane |
Wetted materials | PTFE, PPS, PEEK, Systec AF TM |
Pressure decline | 1.37 mm (Hg/ml/min) |
Maximum pressure stability | 70 psi |
Display | 1 LED |
---|---|
Ambient conditions | temperature range: 4 - 40 °C, 39.2 - 104 °F air humidity: below 90%, non-condensing |
Power supply | 85 - 265 V, 50 - 60 Hz, 20 W |
---|---|
Dimensions | 121 x 138 x 190 mm (W x H x D) |
Weight | 2.3 kg |
Technical data are subject to change without notice.
All trade names and registered trademarks are the property of their respective owners.