Why do I get a pressure peak at valve switching at elevated flowrates?

If you are you are using rotary valves in your HPLC system, there will always be a short time whilst valve switching in which the flow is “in-between” valve positions. The valve drive VU4.1 is developed to minimize this phenomenon by its fast switching times to prevent back-pressure build up. Additionally KNAUER offers a number valves as break-free version.

However, some valves like the multiinjection-valves have longer switching times due to their inherent geometry. When working with these valves, keep in mind that high flowrates* can result in back pressure peaks during the switching process.  

If pressure peaks are observed, you can reduce the flowrate during the injection process and then increase the flowrate after the valve switched to the desired inject position. 

 

*the exact flowrate upon which backpressure peaks can occur depends on used solvents and capillary ID. 

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: