How to prevent the error message MEASUREMENT TIMEOUT?

Cause Solution
The cooling limit in the temperature menu (COOL) is too high. Lower the cooling limit gradually by steps of 0.5 °C to accelerate the cooling process. Please be aware that after the settings have been changed, a new calibration is required.
The cooling device is defective. Check the cooling capacity of the device according to the KNAUER user manual. If the cooling device fails the test, please contact our customer support.
The thermistor is scratched or broken. Please contact our customer support for a replacement.
The ventilation slots are blocked, impeding optimal cooling. Please ensure unhindered ventilation by keeping the slots free.
The initial sample temperature is too high, causing prolonged cooling. Please use samples with a temperature ideally not higher than room temperature.

 

 

 

 

 

 

 

 

 


 

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: