High Sensitive Electrochemical Detector for Liquid Chromatography


single cell control, with DC, pulse and scan mode

With its measurement frequency up to 100 Hz, this electrochemical detector is specially designed for super-fast highly sensitive and selective measurement of oxidisable and reducible substances in (U)HPLC. The AZURA ECD 2.1 comprises of a thermostat-controlled Faraday's cage, accommodating column and flow cell. The AZURA ECD 2.1 unites the three operating modes DC, Pulse and Scan in one instrument. The DC mode covers about 90% of all applications. The pulse mode is important for PAD (Pulsed Amperometric Detection) of e.g. carbohydrates. The scan mode is used to obtain a voltammogram in method optimization. A digital low-pass filter provides an excellent signal-to-noise ratio. For highest sensitivity, the SenCell flow cell is recommended. The correct flow cell can be chosen from a broad variety of flow cells after our advice.

Key Features

Technical Data

Puls mode

Measuring range 10 pA-200 μA in steps of 1, 2 and 5
Filter (cut off) Advanced Digital Filter, 0.5 -0.001 Hz in steps of 1, 2 and 5
Pulse times t1: 100-2000 ms; t2: 0-2000 ms; t3: 0-2000 ms in steps of 10 ms
Data recording 20, 40, 60, 80 and 100 ms

DC mode

Measuring range 10 pA-200 μA in steps of 1, 2 and 5
Filter (cut off) Advanced Digital Filter, 0.5 -0.001 Hz in steps of 1, 2 and 5
Noise < 2 pA with dummy cell

Scan mode

Measuring range 10 pA-200 μA in steps of 1, 2 and 5
Scan speed 1-50 mV/s in steps of 1, 2 and 5
Scan cycle half, complete, continuous

Detection

Detector type Electrochemical detector
Version DC, pulse, scan operating modes
Detection channels 1
Working potential -2.0 V to +2.0 V
Maximum data rate 10 Hz
Autozero triggered via key, TTL, RS-232

Heated zone

Thermostatting oven included
Temperature range from 7°C above ambient temperature to 45°C

Communication

Control RS-232

General

Dimensions 220 x 440 x 440 mm
Weight 14,6 kg without column and flow cell