Are there rules how to configure an assistent ASM 2.1L?

  1. A maximum of two pumps are allowed in each assistant. HPG (high pressure gradient) is not supported.
  2. A maximum of one detector is supported in each assistant.
  3. A pump P 4.1S cannot be placed to the right of a detector UVD 2.1S.
  4. A detector between two pumps is not allowed.
  5. A maximum of two VICI valve drives AWB02 are supported in each assistant.
  6. A maximum of one degasser is supported in each assistant.
  7. An assistant with only one degasser is not allowed.