EN DE

Overview of adjustable Flow Splitters

We offer adjustable Flow Splitters for analytical, semi-preparative and preparative applications.

Analytical post-column Flow Splitters for flow rates of 0.25 - 5.0 ml/min

The default inlet flow for calibration is 1.0 ml/min. However, please always specify your inlet flow (0.25 - 5 ml/min) before order. This way we can assure optimum pressure drop across the splitter, even if the inlet flow differs from the default calibration inlet flow rate.
For orders of spare parts please provide the part number and the serial number of the splitter.

Flow Splitter Split ratio [min.] Split ratio [max.] Inline Filter Inline filter housing Replacement Resistor
A1770-1 50:1 1000:1 A58156 A58159 A58161
A1770-2 15:1 300:1 A58156 A58159 A58162
A1770-3 5:1 100:1 A58156 A58159 A58163
A1770-4 1:1 20:1 A58156 A58159 A58164
Port size: 1/16" OD; UNF 10-32 thread
Max. operating pressure: 350 bar / 5.000 psi
Wetted materials: Stainless Steel, PEEK, Teflon
Dimensions (HxWxD): 12.2 cm x 10.2 cm x 13 cm (4.8" x 4" x 5.1")

Semi-preparative post-column Flow Splitters for flow rates of 5.0 - 40 ml/min

The default inlet flow for calibration is 20.0 ml/min. However, please always specify your inlet flow (5.0 - 40 ml/min) before order. This way we can assure optimum pressure drop across the splitter, even if the inlet flow differs from the default calibration inlet flow rate.
For orders of spare parts please provide the part number and the serial number of the splitter.

Flow Splitter Split ratio [min.] Split ratio [max.] Inline Filter Inline filter housing Replacement Resistor
A5816-2 1000:1 20.000:1 A58152 A58159 A58165
A5816-3 100:1 2.000:1 A58152 A58159 A58166
A5816-4 15:1 300:1 A58152 A58159 A581511
A5816-5 1:1 20:1 A58152 A58159 A58153
Port size: 1/16" OD; UNF 10-32 thread
Max. operating pressure: 350 bar / 5.000 psi
Wetted materials: Stainless Steel, PEEK, Teflon
Dimensions (HxWxD): 12.2 cm x 10,2 cm x 13 cm (4.8" x 4" x 5.1")

Preparative post-column Flow Splitters

The preparative flow splitters are set to a custom split ratio. Therefore, please always specify your inlet flow and the desired split ration before order.
Please note, that the port of the low flow rate outlet has a UNF 10-32 threadi and is for 1/16" OD capillaries. For the ports of the Inlet and the high flow rate outlet we offer also versions with UNF 5/16-24 thread for 1/8" OD capillaries as indicated in the tabel below.
For orders of spare parts please provide the part number and the serial number of the splitter.


Flow Splitter Flow rate Port size
(Inlet & Outlet of high flow)
Inline Filter Inline filter housing Replacement Resistor
A5815-1 40 - 125 ml/min 1/16" OD
UNF 10-32 thread
A58157 A58159 A58167
A5815-2 75 - 200 ml/min 1/16" OD
UNF 10-32 thread
A58158 A58159 A58151
A5815-3 75 - 200 ml/min 1/8" OD
UNF 5/16-24 thread
A58155 A58160 A58151
A5815-4 100 - 1 000 ml/min 1/16" OD
UNF 10-32 thread
A58158 A58159 A58168
A5815-5 100 - 1 000 ml/min 1/8" OD
UNF 5/16-24 thread
A58155 A58160 A58168
Max. operating pressure: 350 bar / 5.000 psi
Wetted materials: Stainless Steel, PEEK, Teflon
Dimensions (HxWxD): 12.2 cm x 10.2 cm x 13 cm (4.8" x 4" x 5.1")

Follow us on: