EN DE

Why the single run or sequence in ClarityChrom does not start after injection?

For a synchronous start of all devices, in ClarityChrom a device must be defined to receive the trigger/start input signal. The device must be defined in the system configuration. In the lower right corner of the configuration window the section "Data Inputs & Outputs" can be found. In "Ext. Start Dig. Input:" select as "Device" the device, which is connected to the injection system (either to manual injection valve or autosampler). The trigger cable must be connected to the Start and Ground pins of the selected device. If an autosampler is used, you also can select the sampler in "Device". In this case, no trigger cable must be connected; the software will detect via the status of the autosampler if an injection is made and start the run. As "Number" always select "1"; otherwise the trigger/start signal will not be detected.

If you use an A/D converter, as the KNAUER IFU 2.1, for the trigger/start input signal, also numbers 2, 3 and 4 are selectable. Select the number that corresponds with the channel number you have connected with the trigger cable. The Start input of the AZURA valve unifier VU 4.1 is not supported in ClarityChrom. If the VU 4.1 or the Virtual Detector is selected as device in "Ext. Start Dig. Input:", runs in a sequence will be started automatically if all devices report a "Ready" status.

In the method setup, on tab "Measurement", the option "External Start/Stop" must be enabled for detecting the trigger/start input signal.

 

Trouble Shooting

1. In ClarityChrom configuration, the correct device must be selected in "Ext. Start Dig. Input:"; "Number" must either be set to "1" or, in case of an A/D converter, to the number, corresponding to the used channel for trigger/start input signal.

2. In the method setup, tab Measurement, the option "External Start/Stop" must be enabled. Additionally, only start options "Start only" and "Down" must be enabled, other options may cause problems.

3. If a sequence is used, the sequence mode "Active" must be selected. This setting can be found in Sequence > Options.

4. In case a trigger cable is connected, the cable must be connected to Start and Ground of the selected device. For some devices, e.g. autosampler Spark Marathon, the cables for Start and Ground are different; make sure the correct cables are connected to Start pin and Ground pin.

5. The Start pin of the AZURA valve unifier VU 4.1 is not supported in ClarityChrom.

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: