To solve this problem, please flush the injector between analyses. If necessary, run strong solvent through column to remove late eluters. Include a final wash step in gradient analyses to remove strongly retained compounds.
Whenever it is possible, use the mobile phase as injector flushing solvent. Also carryover of the flushing solvent can cause ghost peaks.