What can be the reason of a fluctuating detector signal in my FPLC system?

The two most common reasons of a fluctuating UV-signal are the lamp and the flow cell. Make sure that the lamp did not exceed 2000 operating hours. If this is not the case the flow cell could be dirty or there could be air trapped in the flow cell. Therefore, clean the flow cell according to  the manual or flush the flow cell with degassed solvent at higher flow rates. To prevent air bubble formation in the flow cell a back pressure should be applied behind the flow cell. Back pressure can be generated by either tubing with a smaller ID or by installing a back pressure regulator. 

We suggest a back pressure regulator, which is adjustable between 1 - 20 bar (A70087). Most of the times 1 or 2 bar of backpressure is sufficient for a stable UV-signal. Keep in mind not to place the back pressure regulator behind the pH electrode because it is pressure sensitive (pressure limit 5 bar).

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: