How to lower the column back pressure?

If the column back pressure is too high or higher than usual, please check the following suggestions.

a.) Problem in pump, injector, in-line-filter or tubing.

Solution: Disconnect column from system and replace with unions 0.010’’ ID or larger tubing to reconnect the injector to the detector. Run pump at high flow rate (2 - 5 ml/min). If pressure is minimal, see cause 2. If not, isolate cause by systematically eliminating system components. Start with the detector and work back to the pump.

b.) Obstructed column.

Solution: Remove guard column if present and check pressure. Replace guard column if necessary. If column is obstructed, reverse and flush the column while disconnected from the detector. If problem persists, use appropriate restoration procedure. If problem still persists, replace column.

c.) Wrong mobile phase.

Solution: Check mobile phase. Check make-up of mobile phase: Even small changes in composition can affect back pressure.

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: