How to deal with talling peaks?

The problem with talling peaks can occur due to following reasons:


RegularProblem

a) Sample reacting with active sites. Solution: First check column performance with standards. If results for test mix are good, add salts or competing base or acid modifier. b) Wrong column type. Solution: Try another column type (e.g. endcapped column for basic compounds). c) Wrong mobile phase pH. Solution: Adjust pH. For basic compounds, higher pH usually provides more symmetric peaks. d) Wrong injection solvent. Solution: Peaks can tail when sample is injected in stronger solvent than mobile phase. If possible, dissolve sample in mobile phase. e) Small void at column inlet. Solution: Repack top of column with particles of same bonded phase functionality. Continue using column in reversed flow direction.

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: