After switching on the devices, they need some warm up time. Lamps of UV or DAD detectors need to be warmed up for approximately 30 minutes to ensure a stable baseline. The flow cells of RI detectors additionally need to be flushed with the solvent to avoid air bubbles and to adjust the zero glass. Furthermore, the HPLC column also needs time to equilibrate properly.

The equlibration time of a column is dependent on the column dimension/column void volume and used flowrate. A column in a dimension 250 x 4 mm ID has a column void volume of aproximately 2.8 mL. So it takes about 14 minutes at flowrate of 1 mL/min to flush the column with five times the column volume.

Wir trennen Moleküle und verbinden Menschen.

KNAUER ist ein mittelständisches, inhabergeführtes Familienunternehmen in Berlin, das seit 1962 im Dienste der Wissenschaft tätig ist. Wir entwickeln und fertigen Instrumente höchster Qualität für Systeme und Komponenten der Flüssigkeitschromatografie, darunter:

Analytische HPLC / UHPLC
Präparative HPLC
Schnelle Proteinflüssigkeitschromatografie  (FPLC)
Mehrsäulenchromatografie / Simulated Moving Bed (SMB)
Osmometrie

Wir sind weltweiter Partner der Wissenschaft. Heute und in Zukunft. 

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!



München

Christian Ulbricht

Tel: 0170-9275967
E-Mail schreiben
Online Meeting vereinbaren



Frankfurt am Main

Anja Roden

Tel: 0175-7275498
E-Mail schreiben



Leipzig

Stefanie Schulze

Tel: 0171- 3063977
E-Mail schreiben
Online Meeting vereinbaren


   
Österreich & Schweiz 

Marius Segl

Tel: +43 676 3806550
E-Mail schreiben 
Online Meeting vereinbaren

 

Finden Sie Ihre/n
Ansprechpartner/in
in Deutschland ...

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

Besuchen Sie uns auf