Product search:

Back

Sensitive multi-channel UV/VIS detector for HPLC with an ultra fast wavelength scan function and outstanding linearity

Smartline

Smartline UV Detector 2550 VIS

Article number: A5191

Bookmark product

190 nm - 900 nm, depending on lamp typ, without flow cell, including accessories


Details
Detection
Detector type

fast scanning monochromator spectrophotometer, double beam

Version

UV-Vis

Light source

with D2 lamp and halogen lamp

Wavelength range

190–900 nm, depending on lamp type

Spectral bandwidth

< 5 nm

Wavelength accuracy

± 1.5 nm, auto-calibration with deuterium lines, verification with holmium oxide filter

Wavelength precision

0.4 nm RSD

Noise

± 0.5 x 10-5 AU at 254 nm*

Drift

< 0.5 x 10-4 AU/h at 254 nm*

Linearity

0–3 AU*

Wavelength scan

fast scan, depending on set resolution and number of signals

Maximum data rate

80 Hz LAN, 10 Hz RS–232

Flow cell

flow cell excluded

Time constants

0.1; 0.2; 0.5:1; 2; 5; 10 s

Integration time

500 µs–10 ms, determined automatically

Communication
Analog outputs

1.000 AU/V

Analog control input

2 channels, 0 to 5 V, 16-bit

Digital control and output

e.g. for standalone solvent recycling with optional solenoid valves

Control

Digital: RS-232, LAN;

Analog: control voltage

Programming

time-controlled: wavelengths, events, control valves, program links; 19 programs with maximum 99 lines and a wake-up program (lamp)

Technical parameters
GLP

detailed report with lamp recognition, operating hours of system, lamp(s) and step motor, number of lamp ignitions

Conformity

GLP

Display

LCD, 2 lines with 24 characters each

Ambient conditions

temperature range: 4–40 °C; 39.2–104 °F

air humidity: below 90 % humidity (non condensing)

General
Power supply

100–240 V, 50–60 Hz, max. 70 VA

Dimensions

226 x 135 x 410 (W x H x D)

Weight

6.0 kg

note

*according to ASTM E1657-98

Downloads
  Title asc Description asc Languages
Brochure Smartline UV Detektor 2550 Smartline UV Detector 2550 brochure Smartline UV Detector 2550, V7724-1US English
Broschüre Smartline UV Detektor 2550 Smartline UV Detector 2550 brochure Smartline UV Detektor 2550, V7724-1DE Deutsch
Handbuch Smartline UV Detector 2550 / UV-VIS Detektor 2550 Smartline UV Detector 2550 / UV-VIS Detektor 2550 manual Smartline UV Detector 2550 / UV-VIS Detektor 2550, V5190 English
Manual Smartline UV Detektor 2550 / UV-VIS Detektor 2550 Smartline UV Detector 2550 / UV-VIS Detektor 2550 manual Smartline UV Detektor 2550 / UV-VIS Detektor 2550, V5190A Deutsch
Smartline UV Detector 2550 Smartline UV Detector 2550 Sensitive multi-channel UV/VIS detector for HPLC with an ultra fast wavelength scan function and outstanding linearity
Accessories / Spare parts
  Titleasc Article numberasc Descriptionasc
Analytical Flow Cell Analytical flow cell UV A4061V2 10 mm path length, 10 µl volume, 1.1 mm ID, 1/16", stainless steel, for Smartline detector 2520 and 2550
Analytical Flow Cell Analytical flow cell UV A4061 10 mm path length, 10 µl volume, 1.1 mm ID, 1/16", stainless steel, for Smartline detectors 200, 2500, 2520, 2550, 2600, and PLATINblue MW-1
Analytical Flow Cell Analytical flow cell UV A4042 3 mm path length, 2 µl volume, 1 mm ID, 1/16", stainless steel, for Smartline detectors 200, 2500, 2520, 2550, 2600,
Analytical Flow Cell Analytical flow cell UV A4045 3 mm path length, 2 µl volume, 1 mm ID, 1/16", PEEK, for Smartline detectors 200, 2500, 2520, 2600
Preparative Flow Cell Preparative flow cell UV A4066 variable path length, 25 µl volume, 2.3 mm ID, 1/8", stainless steel, for Smartline detectors 200, 2500, 2520, 2550, and 2600
Preparative Flow Cell Preparative flow cell UV A4067 variable path length, 25 µl volume, 2.3 mm ID, 1/8", PEEK, for Smartline detectors 200, 2500, 2520, 2550, and 2600
Preparative Flow Cell Preparative flow cell UV A4068 variable path length, 160 µl volume, 4 mm ID, 1/4" angulate tubing connection, stainless steel, for Smartline detectors 200, 2500, 2520, 2550, and 2600
Preparative Flow Cell Preparative flow cell UV A4069 variable path length, 3 µl volume, 0.8 mm ID, 1/16", stainless steel, for Smartline detectors 200, 2500, 2520, 2550 and 2600
Preparative Flow Cell Preparative flow cell UV A4095 variable path length, 3 µl volume, 0.8 mm ID, 1/16", PEEK, for Smartline detectors 200, 2500, 2520, 2550, and 2600
no photo available Deuterium lamp type HBST A5194 replacement, for Smartline UV-VIS Detector 2550 (A5191), original spare part
no photo available Tungsten halogen lamp A5195 replacement, with lamp housing, for 2-lamp version of Smartline UV Detector 2550, original spare part
Related products
  Titleasc Article numberasc Descriptionasc
Smartline UV Detector 2550 Smartline UV Detector 2550 D2 A5190 190 - 750 nm, without flow cell
Smartline UV Detector 2600 Smartline UV Detector 2600 A5200 PDA technology, 190 – 510 nm, without flow cell, including accessories
Smartline UV Detector 2600 Smartline UV Detector 2600 VIS A5201 PDA technology, 430 - 710 nm, without flow cell, including accessories
Smartline PDA Detector 2850 A5270 UV/VIS/NIR with D2 and halogen lamp, 190 - 1020 nm, without flow cell
Smartline PDA Detector 2850 A5271 UV/VIS, D2 only, 190 - 600 nm, without flow cell

Technical data are subject to change without notice.

Your contact

Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH
Headquarter
Hegauer Weg 38
14163 Berlin
Phone: +49 (0)30 809 727 0
Fax: +49 (0)30 801 50 10
info@knauer.net Contactform