Multi-channel UV spectro-photometer based on diode array technology

Smartline

Smartline UV Detector 2600

Article number: A5200

Bookmark product

The Smartline UV Detector 2600 can simultaneously measure at up to four freely programmable detection channels with high sensitivity and continuously record spectra in a selectable scan-range. The signals of the four channels are additionally distributed on the four analog outputs. With the level and event outputs, it is even possible to set up software-independent solvent recycling by just adding a 3-port solenoid valve. Together with KNAUER ChromGate® or ClarityChrom® software many functions that are typical of a diode array detector are supported including multi-wavelength operation, 2-D and 3-D chromatograms, peak purity, spectra libraries, and more.


Details
Detection
Detector type

diode array spectrophotometer

Version

UV

Number of diodes

256

Pixel pitch

1.25 nm

Detection channels

4

Light source

deuterium

Wavelength range

190–500 nm

Spectral bandwidth

4–25 nm (selectable)

Wavelength accuracy

≤ 1nm (254 nm)

Wavelength precision

0.7 nm RSD

Noise

≤ 1 x 10-5 AU*

Drift

≤ 5 x 10-4 AU/h*

Spectral scan

max. 10 scans per second

Wavelength scan

fast scan, depending on set resolution and number of signals

Time constants

0.1/ 0.2 / 0.5 /1.0 / 2.0 / 5.0 / 10.0 s

Integration time

13–200 ms

Wavelength verification

automatic, with internal holmium oxide filter

Communication
Analog outputs

4 x ± 1 V (scalable in 16 steps), 20 bit

Signal options

inversion, signal ratios (λ-1/λ-2 and vice versa)

Level/event outputs

allows e.g. stand-alone solvent recycling just with an optional solenoid valve

Control

digital: RS-232; analog: remote control strip, key pad

Programming

20 Programs, 9 Links, 1 Wake up

Technical parameters
GLP

detailed trace report with operating hours of total, lamp; number of lamp ignitions, service information

Display

LCD, 2 lines, 24 characters each

Ambient conditions

temperature range: 4–40 °C; 39.2–104 °F

air humidity: below 90 % humidity (non condensing)

General
Power supply

voltage range: 90–260 V, 47– 63 Hz, 70 W

Dimensions

226 x 135 x 412 mm (W x H x D)

Weight

5.7 kg

note

*according to ASTM E1657-94

Downloads
  Title asc Description asc Languages
Brochure Smartline UV Detector 2600 Smartline UV Detector 2600 brochure Smartline UV Detector 2600, V7726-2US English
Manual Smartline UV Detector 2600 Smartline UV Detector 2600 manual Smartline UV Detector 2600, V7605
English
Smartline UV Detector 2600 Smartline UV Detector 2600 Multi-channel UV spectro-photometer based on diode array technology
Accessories / Spare parts
  Titleasc Article numberasc Descriptionasc
Analytical Flow Cell Analytical flow cell UV A4061 10 mm path length, 10 µl volume, 1.1 mm ID, 1/16", stainless steel, for Smartline detectors 200, 2500, 2520, 2550, 2600, and PLATINblue MW-1
Analytical Flow Cell Analytical flow cell UV A4042 3 mm path length, 2 µl volume, 1 mm ID, 1/16", stainless steel, for Smartline detectors 200, 2500, 2520, 2550, 2600,
Analytical Flow Cell Analytical flow cell UV A4045 3 mm path length, 2 µl volume, 1 mm ID, 1/16", PEEK, for Smartline detectors 200, 2500, 2520, 2600
Preparative Flow Cell Preparative flow cell UV A4066 variable path length, 25 µl volume, 2.3 mm ID, 1/8", stainless steel, for Smartline detectors 200, 2500, 2520, 2550, and 2600
Preparative Flow Cell Preparative flow cell UV A4067 variable path length, 25 µl volume, 2.3 mm ID, 1/8", PEEK, for Smartline detectors 200, 2500, 2520, 2550, and 2600
Preparative Flow Cell Preparative flow cell UV A4069 variable path length, 3 µl volume, 0.8 mm ID, 1/16", stainless steel, for Smartline detectors 200, 2500, 2520, 2550 and 2600
Preparative Flow Cell Preparative flow cell UV A4095 variable path length, 3 µl volume, 0.8 mm ID, 1/16", PEEK, for Smartline detectors 200, 2500, 2520, 2550, and 2600
no photo available U-Z View micro flow cell A4091 8 mm path length, 0.14 µl volume, 0.15 mm ID, 1/16", fused silica, for Smartline detectors 200, 2520, 2550, 2600
no photo available U-Z View micro flow cell A4092 8 mm path length, 0.035 µl volume, 0.15 mm ID, 1/16", fused silica, for Smartline detectors 200, 2520, 2550, 2600
no photo available Test cell A4124 normal/normal
no photo available Deuterium lamp A4071 replacement, 35 W, original spare part
no photo available Tungsten-halogen lamp A4073 with base
no photo available RS-232 cable A0895 for connection to a PC, original spare part
Related products
  Titleasc Article numberasc Descriptionasc
Smartline UV Detector 2600 Smartline UV Detector 2600 VIS A5201 PDA technology, 430 - 710 nm, without flow cell, including accessories
Smartline PDA Detector 2850 A5270 UV/VIS/NIR with D2 and halogen lamp, 190 - 1020 nm, without flow cell
Smartline PDA Detector 2850 A5271 UV/VIS, D2 only, 190 - 600 nm, without flow cell

Technical data are subject to change without notice.