UV/VIS/NIR with D2 and halogen lamp, 190 - 1020 nm, without flow cell, including accessories


Details
Detection
Detector type

diode array spectrophotometer

Version

UV/VIS/NIR

Number of diodes

1024

Pixel pitch

0.8 nm

Detection channels

4

Light source

deuterium (D2) and tungsten-halogen

Wavelength range

190–1020 nm

Spectral bandwidth

default: 1 nm, recommended: 8 nm

Wavelength accuracy

≤ 0.5 nm

Wavelength precision

≤ 0.1 nm

Wavelength resolution

0.8 nm/diode

Noise

≤ 1 x 10-5 AU*

Drift

≤ 5 x 10-4 AU/h*

Linearity

0–2.0 AU

Time constants

0.1; 0.2; 0.5; 1; 2; 5; 10 sec (default: 0.1 sec)

Wavelength verification

optional with holmium oxide filter cell

Communication
Analog outputs

2 x ± 10 V (software-scalable)

Level/event outputs

allows e.g. stand-alone solvent recycling just with an optional solenoid valve

Control

LAN

Technical parameters
Ambient conditions

temperature range: 4–40 °C; 39.2–104 °F

air humidity: below 90 % humidity (non condensing)

General
Power supply

50–60 Hz

Dimensions

226 x 190 x 412 mm (W x H x D)

Weight

8.6 kg

Optional accessories

via fiber optics, optional extension available for external use

note

*according to ASTM E1657-94

Downloads
  Title asc Description asc Languages
Brochure Smartline PDA Detector 2850 Smartline PDA Detector 2850 brochure Smartline PDA Detector 2850, V7782-2US English
Broschüre Smartline PDA Detektor 2850 Smartline PDA Detector 2850 brochure Smartline PDA Detector 2850, V7728-2DE Deutsch
Accessories / Spare parts
  Titleasc Article numberasc Descriptionasc
Analytical Flow Cell LWL Analytical flow cell UV A4130 10 mm path length, 10 µl volume, 1.1 mm ID, 1/16", stainless steel, fiber optic for Smartline PDA detector 2850
Analytical Flow Cell LWL Analytical flow cell UV A4131 3 mm path length, 10 µl volume, 1 mm ID, 1/16", stainless steel, fiber optic for Smartline PDA detector 2850
Analytical Flow Cell LWL Analytical flow cell UV A4132 3 mm path length, 2 µl volume, 1 mm ID, 1/16", PEEK, fiber optic for Smartline detector 2850
Preparative Flow Cell LWL Preparative flow cell UV A4133 variable path length and variable volume, 1/8", stainless steel, fiber optic for Smartline detector 2850
Preparative Flow Cell LWL Preparative flow cell UV A4134 variable path length and variable volume, 1/8", PEEK, fiber optic for Smartline detector 2850
Preparative Flow Cell LWL Preparative flow cell UV A4135 variable path length and variable volume, 1/4", stainless steel, fiber optic for Smartline detector 2850
Preparative Flow Cell LWL Preparative flow cell UV A4136 0.5 mm path length, 1/16", stainless steel, fiber optic for Smartline detector 2850
Preparative Flow Cell LWL Preparative flow cell UV A4137 0.5 mm path length, 1/16", PEEK, fiber optic for Smartline detector 2850
no photo available Test cell A4129 holmium oxid filter, double fiber optic connector for PDA 2800
no photo available Test cell A4124 normal/normal
no photo available Deuterium lamp A4447 replacement, 35 W, for PDA 2800
no photo available Tungsten-halogen lamp A4448 7.5 W, with connector for TIDAS II adjusted, for PDA 2800
no photo available Fiber optic cable SMA-SMA A4528 105 mm length, for Smartline PDA Detector 2850 and 2800
no photo available SMA-SMA coupling A4530 fiber optic coupling
no photo available Fiber optic connector, 2 pc. A0740 750 mm, 2x SMA 600/660, polymicro
no photo available Fiber optic connector, 2 pc. A0743 custom made sizes, polymicro, SMA 600/660 2 pieces
Related products
  Titleasc Article numberasc Descriptionasc
Smartline UV Detector 2600 Smartline UV Detector 2600 A5200 PDA technology, 190 – 510 nm, without flow cell, including accessories
Smartline UV Detector 2600 Smartline UV Detector 2600 VIS A5201 PDA technology, 430 - 710 nm, without flow cell, including accessories
Smartline PDA Detector 2850 A5271 UV/VIS, D2 only, 190 - 600 nm, without flow cell