analytical HPLC autosampler with sample tray cooling
Article number: A50071The Smartline Autosampler Knauer Optimas was developed for optimal automation of routine injections.
With a choice of carousel trays, it provides sample vial flexibility and with the proven PASA™ injection concept, reliable and accurate sample handling is guaranteed. The autosampler also features three distinct injection modes, “full loop filling” (highest precision and reproducibility), “partial loop filling” (variable volumes, e.g. for dilution series) and “µl pickup” (loss-free injection of extremely small sample volumes), allowing the user to optimize sample usage.
The autosampler is equipped with a sample tray containing 84 positions for 1.5 ml vials and 3 positions for 10 ml vials by default. Precolumn derivatization is supported. For all injection modes, sample injection with pressure-assisted sample aspiration (PASA™) is selectable. This technique reduces the risk of bubbles and provides excellent accuracy. The needle does not move to an extra port for injection, and thus prevents contamination. Complete integration into KNAUER’s ChromGate® (based on the EZChrom Elite™ kernel) and ClarityChrom® chromatography data software make the autosampler easy to control. Alternatively, the autosampler can be used in standalone mode.
With sample tray cooling (4 - 22°C).
| Sample injection | |||||||
|---|---|---|---|---|---|---|---|
| Sample capacity | standard tray: for 84 vials x 1.5 ml + 3 vials x 10 ml |
||||||
| Analytical performance | |||||||
| Injection volumes | 1– 5 000 μl (preparative option: 10 000 μl when software-controlled) |
||||||
| Sample loop | 100 μl standard; 10 ml for preparative option |
||||||
| Dispenser syringe | 250 μl standard; 1 000 μl optional; 2500 μl for preparative option |
||||||
| Vial detection | missing vial detection by sensor |
||||||
| Headspace pressure | built-in compressor |
||||||
| Switching time inj. valve | < 100 ms |
||||||
| Piercing precision needle | ± 0.6 mm |
||||||
| Wash solvent | external wash solvent bottle |
||||||
| Wetted materials | stainless steel (SS316), PTFE, Tefzel®, Vespel®, glass, Teflon®. For biocompatible version: PEEK |
||||||
| Injection cycle time | 73 sec for 10 μl partial loopfill including 250 μl needle wash |
||||||
| Analytical performance | |||||||
| Injection modes | full loop, partial loopfill and μl-pickup mode; PASA™ (pressure-assisted sample aspiration) |
||||||
| Injection volumes |
|
||||||
| Injections per vial | max. 9 injections (unlimited when controlled by software; in sequence mode: max. 9 injections) |
||||||
| Injection precision |
|
||||||
| Analysis time | max. 9 h, 59 min, 59 s |
||||||
| Sample carryover | < 0.05 % (prep. < 0.1 %) with programmable needle wash |
||||||
| Communication | |||||||
| Outputs | 2 inject marker (relay, TTL), 2 auxiliary (relay), alarm (relay) |
||||||
| Inputs | next injection, freeze, stop (all TTL) |
||||||
| Control | RS-232, standalone |
||||||
| General | |||||||
| Power supply | 115 – 230 Volt AC +10 % / -22 %; 50 – 60 Hz; 200 VA |
||||||
| Dimensions | 300 mm x 500 mm x 340 mm (W x H x D) |
||||||
| Weight | 21 kg |
||||||
|
Title
|
Description
|
Languages | |
|---|---|---|---|
|
Smartline Autosampler KNAUER Optimas brochure | Smartline Autosampler KNAUER Optimas, V7736US | English |
|
Title |
Article number |
Description |
|
|---|---|---|---|
| Preparative option kit for Autosampler KNAUER Optimas | A50072 | tray for 24 x 10 ml vials, large bore injection valve, 2.5 ml syringe, 10 ml sample loop | |
| Biocompatible option kit for Autosampler KNAUER Optimas | A50073 | PEEK injection valve, biocompatible needle, 100 μl PEEK sample loop | |
| Tray for Autosampler KNAUER Optimas | A50074 | 96 positions for 1,5 ml vials | |
| Tray for Autosampler KNAUER Optimas | A50075 | 24 positions for 10 ml vials | |
| Wash bracket | A50076 | for 2 x 250 ml wash bottles (including wash bottles) |