Product search:

Back

1 Cortisone m/z = 361, 2 Corticosterone m/z = 347, 3 Testosterone m/z = 289, 4 Deoxycorticosterone m/z = 331, 5 Norgestrel m/z = 313, 6 Progesterone m/z = 315

Separation of Steroids with MS detection

Article number: VPH0043J

Bookmark product

1 Cortisone, 2 Corticosterone, 3 Testosterone, 4 Desoxycorticosterone, 5 Norgestrel, 6 Progesterone


Details
Method

UHPLC

Eluent

A: H2O + 0.1 % FAB: ACN + 0.1 % FA

Gradient

0.0 – 1.0 min 40 % B1.0 – 1.5 min 40 - 90 % B1.5 – 2.0 min 90 % Bs

Flow rate

0.4 ml/min

Injection volume

5 µl

Temperature

40 °C

Injection volume

5 µl

Temperature

40 °C

Detection

MSQ Plus, ESI pos. mode,Needle 3.5 kV, coneV 20 V,Probe temperature 350 °C,full scan m/z = 270 – 370

Substances

Cortisone. Corticosterone, Testosterone, Deoxycorticosterone, Norgestrel, Progesterone

Chromatogram

1 Cortisone m/z = 3612 Corticosterone m/z = 3473 Testosterone m/z = 2894 Deoxycorticosterone m/z = 3315 Norgestrel m/z = 3136 Progesterone m/z = 315

Downloads
  Title asc Description asc Languages
Chromatotgramm VPH0043J 1 Cortisone m/z = 361, 2 Corticosterone m/z = 347, 3 Testosterone m/z = 289, 4 Deoxycorticosterone m/z = 331, 5 Norgestrel m/z = 313, 6 Progesterone m/z = 315
UHPLC-MS method for the quantification of steroid hormones Cortisone, Corticosterone, Deoxycorticosterone, Norgestrel, Progesterone, Testosterone English
Related products
  Titleasc Article numberasc Descriptionasc
Vertex Plus Column 50 x 2 mm BlueOrchid 175-1.8 C18 05BI181BOE Vertex Plus Column 50 x 2 mm BlueOrchid 175-1.8 C18

Technical data are subject to change without notice.

Your contact

Columns and applications department
Phone: +49 (0)30 809 727-0
Fax: +49 (0)30 801 50 10
applications@knauer.net Contactform

Convert your methods from HPLC to UHPLC