Product search:

Back

Peptide mixture; (customer sample)

Separation of peptides

Article number: VBS0002J

Bookmark product


Details
Method

HPLC/UHPLC

Eluent

A: Water (0.1 % TFA (Trifluoroacetic acid))

B: 60 % Acetonitrile (0.1 % TFA (Trifluoroacetic acid))

 

Gradient

0.0 – 10.0 min 0 – 100 % B

10.0 – 13.0 min 100 % B

 

Flow rate

0.5 ml/min

Mode

RP-Mode

Injection volume

5 µl

Pressure

540 bar (7830 psi)

Temperature

40 °C

Detection

PDA-1, 210 nm (50 Hz, 0.01 s)

10 mm, 2 µl fl ow cell

 

Chromatogram

Peptide mixture

(customer sample)

 

Downloads
  Title asc Description asc Languages
Chromatogram VBS0002J Chromatogram VBS0002J Peptide mixture; (customer sample)
Related products
  Titleasc Article numberasc Descriptionasc
Vertex Plus Column 100 x 2 mm BlueOrchid 175-1.8 C18A 10BI184BOE Vertex Plus Column 100 x 2 mm BlueOrchid 175-1.8 C18A