1. Bisphenol F; 2. Bisphenol A

Separation of Bisphenol A and Bisphenol F

Article number: VFD0014J

Bookmark product


Details
Method

HPLC

Eluent

A: Water

B: Acetonitrile

Gradient

0 – 3 min 50 % B

3 – 3.1 min 50 % – 95 % B

3.1 – 10 min 95 % B

10 – 10.1 min 95 % – 50 % B

10.1 – 22 min 50 % B

 

Flow rate

0.6 ml/min

Mode

RP Mode

Injection volume

5 µl

Temperature

30 °C

Detection

UV at 227 nm

Substances

Bisphenol A, Bisphenol F

Key words

Bisphenol, PVC/Polycarbonate additives, Plastics, Baby bottles