Product search:

Back

Retention time standard; 1. Cyclohexane; 2. Phenyldodecane; 3. 1,2 Dimethylbenzene; 4. Hexamethylbenzene; 5. Naphthalene; 6. Dibenzothiophene; 7. 9-Methylanthracene; Calibration standard; 8. 1,2 Dimethylbenzene; 9. Fluorene; 10. Phenanthrene; Sample; 11. MAH 20.14% [m/m]; 12. DAH 2.33% [m/m]; 13. T+AH 0.11% [m/m]; Total aromatic hydrocarbens:; 22.59 % [m/m]

Determination of aromatic Hydrocarbon types in middle distillates acc. to prEN 12916:2006

Article number: VCH0002J

Bookmark product


Details
Method

HPLC

Sample

Dilution with n-heptane

Eluent

n-Heptane (dried with molecular sieve)

Gradient

isocratic

Flow rate

0.5 ml/min

Mode

NP Mode

Injection volume

10 µl

Temperature

25 °C

Detection

Refractive Index (RI)

Substances

MAH: phenyldodecane; 1,2 dimethylbenzene; hexamethylbenzene

DAH: naphthalene, fluorene

T-AH: dibenzothiphene; phenanthrene, 9-methylanthracene

 

Key words

MAH (Mono-aromatic compounds), DAH (Di-aromatic compounds), T+AH (Tri+-aromatic compounds), middle distillates

Chromatogram

Retention time standard

1. Cyclohexane

2. Phenyldodecane

3. 1,2 Dimethylbenzene

4. Hexamethylbenzene

5. Naphthalene

6. Dibenzothiophene

7. 9-Methylanthracene

Calibration standard

8. 1,2 Dimethylbenzene

9. Fluorene

10. Phenanthrene

Sample

11. MAH 20.14% [m/m]

12. DAH 2.33% [m/m]

13. T+AH 0.11% [m/m]

Total aromatic hydrocarbens:

22.59 % [m/m]

 

Related products
  Titleasc Article numberasc Descriptionasc
Column 250 x 4 mm UltraSep 100-5 ES AP special 25DE450USJ Column 250 x 4 mm UltraSep 100-5 ES AP special

Technical data are subject to change without notice.

Your contact

Columns and applications department
Phone: +49 (0)30 809 727-0
Fax: +49 (0)30 801 50 10
applications@knauer.net Contactform

Convert your methods from HPLC to UHPLC