1. Hydroxypyrene

Biomonitoring of Hydroxypyrene

Article number: VEV0017J

Bookmark product


Details
Method

HPLC

Eluent

A: Acetonitrile

B: Water

 

Gradient

0 – 28 min 60% – 100% A

5 min hold, equilibrate column 10 min with 60% A for next injection

 

Flow rate

1.0 ml/min

Mode

RP Mode

Injection volume

10 µl

Temperature

30 °C

Detection

Fluorescence: excitation 242 nm, emission 388 nm

Substances

Hydroxypyrene (Biomarker of PAH contamination)

Key words

Hydroxypyrene , Biomarker , PAH Contamination

Chromatogram

1. Hydroxypyrene

 

Downloads
  Title asc Description asc Languages
Chromatogram VEV0017J Chromatogram VEV0017J 1. Hydroxypyrene
Related products
  Titleasc Article numberasc Descriptionasc
VertexPlus Column 250 x 4 mm Eurospher 100-5 C18 25DE181ESJ VertexPlus Column 250 x 4 mm Eurospher 100-5 C18